Shutting Down the System
Shutdown Cleaning
- Discard any remainder of the sample in the nozzle and run "Wash Flush Strong" at least twice.
- Run "Wash Tray Station" to wash the nozzle with 0.5% SciClean 8. The SciClean solution must have been prepared the same day. The recommended concentration is 0.5%, and can be obtained by adding 50 μL of SciClean 8 to 1 mL of filtered water. The solution must be discarded at the end of the day.
- Run "Wash Flush Strong" again.
- Close the software. Please be patient, the software may take some time to close as it writes the setting and log files on the disk.
- Once the software is closed, shut down the computer.
- Turn off the power bar.
- Discard system fluid left in the system fluid bottle. You can recycle it by discarding it in the "Wash Water" carboy.
- Discard the contents of the waste carboy if applicable.