Documentation:Standard Operating Procedures for Vacuum Oven Dryer
- Please obtain proper training from the FNH teaching lab technician Imelda Cheung (Imelda.Cheung@ubc.ca) prior to using the vacuum oven dryer.
- Turn on the power (green button) and the temperature should be set at 70oC (setting 3).
- Place pre-weighed samples in the main chamber.
- Samples should always be handled by tongs as fingerprints will add weights to the foil dish.
- Close the chamber door tightly.
- The Vacuum Valve should remain fully open at all times and need not be adjusted.
- The Vent Valve should be fully closed.
- Manually plug in the power cord (yellow) to start the vacuum pump.
- The vacuum gauge should slowly reach ~ 25 in. Hg.
- When drying is complete, unplug the vacuum pump power cord.
- Release the vacuum by slowly opening the Vent Valve until vacuum gauge reads zero.
- Releasing vacuum quickly may cause loss of samples!
- Remove samples and place in a dessicator to cool before weighing.
- Turn off the power.
- To reserve a time to utilize the vacuum oven dryer, please visit https://www.qreserve.com/r/resource/QR00182640