Documentation:Standard Operating Procedures for VWR pH Meter

From UBC Wiki
VWR pH meter
  1. Please obtain proper training from the FNH teaching lab technician Imelda Cheung (Imelda.Cheung@ubc.ca) prior to using the pH meter.
  2. Turn on pH meter by pressing “Ф”.
  3. Press “Cal” to access the clear pH standardization menu.
  4. Press “Clear” to display clear buffer icon.
  5. Press “Clear” again to clear all existing buffers.
  6. Remove pH probe carefully from storage bottle and turn the probe to “ON” position. Only if applicable. Some pH probes may not have any opening and can be used directly..
  7. Rinse with ddH2O and gently dry with kimwipes.
    As much as possible, keep the pH probe steady on a clamp and move the buffers, samples or solutions to under the pH probe for measurement.
  8. Immerse the probe in pH 4 buffer and press “Read” and wait for value to stabilize.
    Autoread function is assumed to be on, as indicated by the eye icon. When the eye is flashing, it’s still stabilizing; when the eye icon stops flashing, the reading is stabilized and entered. Further changes to pH will not be recorded until you press the read button again.
  9. Repeat steps 7 – 8 for pH 7 and 10 buffers.
  10. The percent slope displayed at the end of calibration should be within the range of 90 – 110%.
  11. When the calibration is complete, press “Cal” to return to reading mode. The electrode is now ready for use.
  12. When pH measurement is complete, turn off the pH meter by pressing “Ф”.
  13. Carefully rinse the pH probe with ddH2O and gently dry with kimwipes.
  14. Put the probe back into the storage bottle.
    Some probes may not have a storage bottle but are stored in a falcon tube filled with buffer solution. Remember to parafilm the tube to avoid drying out of probes.
To reserve a time to utilize the pH meters, please visit https://www.qreserve.com/r/resource/QR00182913