Documentation:Standard Operating Procedures for Sorvall RC-5B Refrigerated Superspeed Centrifuge
- Please obtain proper training from the FNH teaching lab technician Imelda Cheung (Imelda.Cheung@ubc.ca) prior to using the Sorvall RC-5B Refrigerated Superspeed CentrifugeSorvall RC-5B Refrigerated Superspeed Centrifuge.
- A video https://www.youtube.com/watch?v=khHhu70CBYQ is also provided for review purpose and can be viewed as needed.
- Turn on the power.
- To set desired temperature, adjust the blue needle (lower limit) using the outside knob and the red needle (upper limit) using the inside knob.
- If using refrigerated temperature, close the chamber door and let cool for at least 15 min.
- To open chamber door, press the “Door” button at the front of the instrument while lifting the door at the same time.
- Properly position the SS-34 rotor in the centrifuge.
- The pins on the rotor should NEVER align with the pins on the spindle. They should be best at perpendicular position to each other.
- Balance the weight of samples + tubes to 0.001 g and place in the rotor. Even number of samples is required and should be placed across from each other.
- Never parafilm samples and put in the centrifuge, the parafilm may get stuck during the spin and you will not be able to retrieve the tubes easily.
- Tighten the rotor lid by turning the big disc, then further tighten using the small disc.
- Close the chamber door.
- Set the appropriate speed in rpm using the control dial (gravity conversion can be found on the chart posted on the chamber door).
- The center part of the chart contains all the universal x g values while the top row describes the types of rotor while the left column displays the rpm for the centrifuge setting. Maximum speed for SS-34 is 15,000 rpm!!
- Set the time desired for the run using the time control dial.
- If there is imbalance detected, you should see the light goes on. It is very important not to begin centrifugation when imbalance is detected as this will cause an explosion.
- Push the start button to begin centrifugation.
- When the run is complete, the door light will be on and the speed display should return to zero.
- Open the chamber door, open the rotor lid and retrieve samples.
- Wipe the interior of the centrifuge with paper towel and rinse the rotor gently with warm water in case of any spills.
- Turn off power.
- Sign the log book.
- To reserve a time to utilize the centrifuge, please visit https://www.qreserve.com/r/resource/QR00182797