Documentation:Standard Operating Procedures for Autoclave
Appearance

- Please obtain proper training from the FNH teaching lab technician Imelda Cheung (Imelda.Cheung@ubc.ca) prior to using the autoclave.
- A video https://www.youtube.com/watch?v=dgvOc2PvzBc is also provided for review purpose and can be viewed as needed.
- Sample/waste handling prior to autoclave:
- Biohazardous wastes should be placed in clear autoclave bag, with an autoclave indicator tape placed in a visible location. The bag should be loosely open to allow steams in and be placed in a metal tray.
- Samples should also be placed in metal tray in case of spills.
- Before operating the steam generator, it's a good practice to flush the steam boiler before use. Please see https://www.youtube.com/watch?v=LkPz8CT2n_Q
- Turn on the Steam Generator (Grey box).
- Turn on the power switch on the wall.
- Inspect the watch glass and ensure there is water, then turn on the Boiler.
- Wait for approximately 15 min until the steam generator gauge reads above 60 psi.
- On the autoclave, turn the valve to “LOAD” position.
- Wait for the jacket pressure to reach over 16 psi. This should take roughly 15 – 20 min.
- If jacket pressure does not reach desired level (i.e. 16 psi) after 30 min, close the chamber door tightly, run the cycle (LOAD --> STERILIZE --> VENT/DRY --> OFF) to vent out the water. For each cycle the chamber pressure MUST return to 0 psi before turning to OFF.
- For each autoclave run, you must sign the logbook and place a chemical indicator strip along with your load to ensure proper temperature is reached in the autoclave.
- Put the wastes or samples in a metal tray into the chamber and close the door tightly.
- Turn the valve to “STERILIZE” position and wait for the chamber pressure to reach 15 psi before manual timing
- Biohazardous wastes - 60 min
- Samples - 15 min
- When the sterilization time is up,
- Biohazardous wastes - turn the valve to “VENT/DRY”
- Equipment and glassware - turn the valve to “VENT/DRY”
- Samples/media – turn the valve to “LIQUID COOL”
- It is important to use “LIQUID COOL” for any liquid containing sample preparation as the “VENT/DRY” function will drop the pressure too quickly and lead to spilling out of solutions from the flasks/bottles.
- Wait until the chamber pressure drops to 0 psi.
- Turn the valve to “OFF” and open chamber door to remove sample/waste.
- If you have further samples/wastes, return to step 8 and repeat. Otherwise, proceed to turn off the Boiler, then the power switch on the wall, and finally the Steam Generator (Grey box).
- At this point, please make sure you have signed the log book or you may not be allowed to utilize the autoclave in the future.
- To reserve a time to utilize the autoclave, please visit https://www.qreserve.com/r/resource/QR00181335